Separation type: Liquid Chromatography Mixed-mode
High Performance Liquid Chromatography (HPLC) Method for Analysis of Ivermectin.
Ivermectin is an anti-parasitic drug that is often used to treat or prevent parasites in animals and humans. Ivermectin can be retained and separated on the Primesep 100 mixed-mode column using an isocratic analytical method with a simple mobile phase of water, acetonitrile (MeCN), and sulfuric acid (H2SO4) buffer. This analysis method can be UV detected at 250 nm.
Column | Primesep 100, 4.6×150 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O – 60/40% |
Buffer | H2SO4 – 0.1% |
Flow Rate | 1.0 ml/min |
Detection | UV, 200 nm |
Class of Compounds |
Drug, Antibiotics |
Analyzing Compounds | Ivermectin |
Application Column
Primesep 100
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.