Fast Blue B Salt

CAS Number14263-94-6
Molecular FormulaC14H12Cl4N4O2Zn
Molecular Weight475.5
InChI KeyGPPKNJIWDULNQH-UHFFFAOYSA-J
Synonyms

    Applications:

    HPLC Method for Analysis of Fast Blue B Salt on BIST B+ Column

    December 7, 2022

    HPLC Method for Analysis of Fast Blue B Salt on BIST B+ by SIELC Technologies.

    HPLC Method for Analysis of Fast Blue on BIST B+ Colum
    HPLC Method for Analysis of Fast Blue B Salt on BIST B+ by SIELC Technologies.

    Fast Blue B Salt is a multicharged basic dye with the chemical formula
    C14H12Cl4N4O2Zn
    . It has a a variety of applications in chromatographic testing, including in thin layer chromatography, lipase activity testing. and in testing acid phosphatase in Clostridium perfringens.

    Using SIELC’s newly introduced BIST™ method, Fast Blue B Salt can be retained on a positively-charged anion-exchange BIST™ B+ column. 

    There are two keys to this retention method: 1) a multi-charged, negative buffer, such as Sulfuric acid (H2SO4), which acts as a bridge, linking the positively-charged analytes to the positively-charged column surface and 2) a mobile phase consisting of a majority of organic solvent (such as MeCN) to minimize the formation of a solvation layer around the charged analytes. Utilizing a step gradient to switch to a completely aqueous MP after 2 minutes allows for retention to occur while also preventing the method from being too long. Using this new and unique analysis method, Fast Blue B Salt can be separated, retained, and UV detected at 320 nm.


    Condition

    ColumnBIST B+, 4.6 x 150 mm, 5 µm, 100 A, dual ended
    Mobile PhaseGradient MeCN
    BufferH3PO4 – 0.2%
    Flow Rate1.0 ml/min
    DetectionUV 320 nm
    Peak Retention Time6.01 min

    Description

    Class of CompoundsDye
    Analyzing CompoundsFast Blue B Salt

    Application Column

    BIST B+

    Column Diameter: 4.6 mm
    Column Length: 150 mm
    Particle Size: 5 µm
    Pore Size: 100 A
    Column options: dual ended

    Add to cart
    Application Analytes:
    Fast Blue B Salt

    Application Detection:
    UV Detection
    SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.