Xylitol is a polyalcohol which is used as a sweetener for people with diabetes. Xylitol is naturally found in the fibers of fruits and vegetables. It is a very highly polar, non-ionizable molecule. Xylitol can be well retained on the Primesep S HILIC column. It has a relatively high water concentration in the mobile phase that helps to dissolve xylitol at high acetonitrile HILIC applications. Polar surface on the Primesep S column provides well controlled retention time and can be used as a robust method for analysis of various sugar-lie molecules.
Column | Primesep S, 4.6×150 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O – 80/20% |
Buffer | AmFm pH 3.0- 10 mM |
Flow Rate | 1.0 ml/min |
Detection | ELSD |
Class of Compounds |
Sugar, Hydrophilic, Ionizable |
Analyzing Compounds | Xylitol |
Application Column
Primesep S
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.