Kynurenine is an important biological compound which carries out a number of biological functions including dilating blood vessles and regulating immune response. Kynurenine was retained on Legacy L1 which containes embedded C18 groups. Comparisons between this chromatagram and from Phenomenex columns are available upon request.
Column | Legacy L1, 4.6×250 mm, 5 µm, 100A |
Mobile Phase | Gradient MeCN – 0-80%, 20 min |
Buffer | No |
Flow Rate | 1.0 ml/min |
Detection | UV, 225 nm |
Class of Compounds |
Acid, Hydrophilic, Ionizable |
Analyzing Compounds | Kynurenine |
&
Application Column
Legacy L1
SIELC's family of Legacy columns is based on the United States Pharmacopeia's (USP) published chromatographic methods and procedures. Numerous brands have columns used in USP reference standards and methods. USP has created various designations to group together columns with similar types of packing and properties in the solid phase. SIELC's Legacy columns adhere to these strict requirements and properties, allowing you to easily replace older columns that are no longer available without needing to significantly modify your method or SOPs.