CAS Number | 1491-59-4 |
---|---|
Molecular Formula | C16H24N2O |
Molecular Weight | 260.382 |
InChI Key | WYWIFABBXFUGLM-UHFFFAOYSA-N |
LogP | 4.01 |
Synonyms |
|
Applications:
HPLC Separation of EDTA, Oxymetazoline and Benzyl Alcohol in Nasal Spray
September 4, 2020
Separation type: Liquid Chromatography Mixed-mode
High Performance Liquid Chromatography (HPLC) Method for Analysis of EDTA, Oxymetazoline and Benzyl Alcohol
Benzyl alcohol is found in over-the-counter medications, topical creams, lotions, shampoos, and facial cleansers as an antibacterial, preservative, and/or fungicide. Benzyl alcohol is a colorless liquid with a mild pleasant aromatic scent. Benzyl alcohol is found in many essential oils including jasmine, hyacinth, neroli, rose, and ylang-ylang. Oxymetazoline is a decongestant that shrinks blood vessels in the nasal passages. All active compounds of nasal spray can be separated in HPLC on a Newcrom BH mixed-mode column. The analytical method uses acetonitrile (ACN) gradient and water with sulfuric acid (H2SO4) as buffer and UV detected at 260nm.
Column | Newcrom BH, 4.6×150 mm, 100A |
Mobile Phase | Gradient MeCN – 2- 70%, 10 min |
Buffer | H2SO4 – 0.1% |
Flow Rate | 1.0 ml/min |
Detection | UV 260nm |
Class of Compounds | Acid, Hydrophilic, Preservatives |
Analyzing Compounds | EDTA, Oxymetazoline, Benzyl Alcohol |
Application Column
Newcrom BH
The Newcrom columns are a family of reverse-phase-based columns. Newcrom A, AH, B, and BH are all mixed-mode columns with either positive or negative ion-pairing groups attached to either short (25 Å) or long (100 Å) ligand chains. Newcrom R1 is a special reverse-phase column with low silanol activity.
Select optionsEDTA (Ethylenediaminetetraacetic Acid)
Oxymetazoline
Oxymetazoline hydrochloride
Improved Peak Shape of a Basic Compound on Primesep B2 Column
June 10, 2004
Primesep B2 shows superior efficiency, symmetry, and peak shape for a hydrophilic basic compound compared to two leading reversed-phase C18 columns. Primesep B2 contains an embedded basic functional group that shields the underlying silica by ionic repulsion with the analyte. On reversed-phase columns exposed active silanols provide unintended polar interactions with basic analytes that lead to tailing. The HPLC separation uses a simple mobile phase of water, acetonitrile (MeCN, ACN) and trifluoracetic acid (TFA) with UV detection at 250 nm.
Column | Primesep B2, 4.6×150 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O |
Buffer | TFA |
Flow Rate | 1.0 ml/min |
Detection | UV, 250 nm |
Class of Compounds |
Drug, Acid, Hydrophilic, Ionizable, Vitamin, Supplements, Amino acid |
Analyzing Compounds | Oxymetazoline |
Application Column
Primesep B2
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.
Select options