Sulcatone is a ketone used as a mosquito attractant. It is a volatile oil component found in
citrus. It can be retained on a Newcrom R1reverse-phase column with a mobile phase of acetonitrile (ACN) and water with sulfuric acid (H2SO4) buffer. The analytical method is MS-compatible and can also use UV detection at 200nm.
Column | Newcrom R1, 4.6×150 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O – 50/50% |
Buffer | H2SO4 – 0.1 % |
Flow Rate | 1.0 ml/min |
Detection | UV, 200 nm |
Class of Compounds |
Hydrophobic, Mosquito attractant |
Analyzing Compounds | 6-methyl-5-hepten-2-one |
Application Column
Newcrom R1
The Newcrom columns are a family of reverse-phase-based columns. Newcrom A, AH, B, and BH are all mixed-mode columns with either positive or negative ion-pairing groups attached to either short (25 Å) or long (100 Å) ligand chains. Newcrom R1 is a special reverse-phase column with low silanol activity.