HPLC Separation of Epinephrine on Legacy L1 Column

Epinephrine was analyzed according to the US Pharmacopeia method on traditional reversed-phase HPLC columns, classified as a L1 type of column according to USP classification. Legacy L1 column can be used for the analysis of drug substances in various formulations. Legacy L1 can be used in new HPLC methods as well as in validated methods which require experiments based on USP manuscript. The mobile phase can be modified to accommodate other detection techniques, like LC/MS.

Condition

Column Legacy L1, 4.6x150 mm, 5 µm, 100A
Mobile Phase MeOH - 15%
Buffer Sodium 1 -Octane sulfonate/EDTA
Flow Rate 1.0 ml/min
Detection UV, 280 nm
 

Description

Class of Compounds Drug, Neurotransmitter, Hydrophobic, Ionizable
Analyzing Compounds Epinephrine
 
SIELC Technologies usually develops more than one method for each compound on our website. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound of interest.
Application Analytes:
Epinephrine

Application Detection:
UV Detection